Operate and Control
- Power Supply : OXFORD IPS 120-10 (Max.12 Amp @10Volt)
- Software Remote-control: IPS_120-10_v03.vi
- ! Change of Range when increasing the current at ~4.3A, when decreasing at ~3.4A! Don't get scared by the noise :)
- Persistent current switch:
- Upper magnet - BOX VII, X9:1/4 ,RHeater(4K)=135.
- IHeater=55mA. The TVL-10 is used as current supply.
- The region of the experimental platform between the first nuclear stage and the dilution refrigerator is compensated to 5mT when both magnets are at their full field.
Magnetisation and Demagnetisation Procedure
- After ramping the magnet the first nuclear stage must be precooled again. The temperature dependence varies as T~t-1/3.
- The first nuclear stage was demagnetized by reducing the field in steps by a factor of 2 to a final field of few mT.
- Sweep Rate B>0.5T : R=2T/h.
- Sweep Rate B<0.5T : decrease by a factor of 2 proportional to the field.
Upper Magnet UMM
- Connection on window side of cryo. Power-cable are marked red (+) and blue (-).
- Bmax = 8T, diameter 100mm.
- I/B-Conversion: B(I) = 0.099 T/A * I[A] , ( 80.68 A = 8 T )
- Sweep Rate: 1 T/h = 10 A/h = 0.167 A/min
Lower Magnet LMM
- LMM NOT YET TESTED SO FAR!
- Connection on door side of cryo. Power-cable are marked green (+) and white (-).
- Bmax = 9T, diameter 50mm.
- I/B-Conversion: B(I) = 0.115 T/A * I[A] , ( 78.4 A = 9 T )
- Sweep Rate: 1 T/h = 8.7 A/h = 0.145 A/min
Connection
- To Power supply: M6 (#13 box wrench, #6 Allen key).
- To Upper Magnet: Connection on window side. Power-cable are marked red (+) and blue (-).
- To lower magnet: Connection on door side of the cryostat. Power-cable are marked green (+) and white (-)
- AMP-Plug (connected witch BOX VII): Control of persistent current switch, read-out of two Pt-100, heater