MAGNET
From KIP Wiki
⧼kip-jumptonavigation⧽⧼kip-jumptosearch⧽
THE 9T MAGNET
... is working but this page is under construction :)
In the picture a threaded rod is attracted by the magnet at 50A.
Operate the Magnet
- Power Supply : OXFORD IPS 120-10 (Max.12 Amp @10Volt)
- Software Remote-control: IPS_120-10_v03.vi
- ! Change of Range when increasing the current at ~4.3A, when decreasing at ~3.4A! Don't get scared by the noise :)
- Upper Magnet, Connection on window side of cryo. Power-cable are marked red (+) and blue (-).
- Persistent current switch:
- Upper magnet - BOX VII, X9:1/4 ,RHeater(4K)=135.
- IHeater=55mA. The TVL-10 is used as current supply.
Magnetisation and Demagnetisation Procedure
- After ramping the magnet the magnetized first nuclear stage must be precooled. The temperature dependence varies as T~t-1/3.
- The first nuclear stage was demagnetized by reducing the field in steps by a factor of 2 to a final field of few mT.
- Sweep Rate B>0.5T : R=2T/h.
- Sweep Rate B<0.5T : decrease by a factor of 2 proportional to the field.
Upper Magnet UMM
- I/B-Conversion: 10A=1T (? not yet clarified, date: 14.Dec 2011)
Lower Magnet LMM
- I/B-Conversion: 0.8T=6.96A -> B (I[A]) = 0.115 T/A *I[A]
Connection
- To Power supply: M6 (13Rigschlüssle, 6erInbus).
- To Upper Magnet: Connection on window side.
- To lower magnet: Connection on door side of the cryostat.