** Upper magnet - BOX VII, X9:1/4 ,R<sub>Heater</sub>(4K)=135{{O}}.
** Upper magnet - BOX VII, X9:1/4 ,R<sub>Heater</sub>(4K)=135{{O}}.
** I<sub>Heater</sub>=55mA. The TVL-10 is used as current supply.
** I<sub>Heater</sub>=55mA. The TVL-10 is used as current supply.



{{clear}}
{{clear}}


== Magnetisation and Demagnetisation Procedure ==
== Magnetisation and Demagnetisation Procedure ==

* After ramping the magnet the magnetized first nuclear stage must be precooled. The temperature dependence varies as T~t<sup>-1/3</sup>.
* After ramping the magnet the magnetized first nuclear stage must be precooled. The temperature dependence varies as T~t<sup>-1/3</sup>.
* The first nuclear stage was demagnetized by reducing the field in steps by a factor of 2 to a final field of few mT.
* The first nuclear stage was demagnetized by reducing the field in steps by a factor of 2 to a final field of few mT.
* Sweep Rate B>0.5T : R=2T/h.
* Sweep Rate B>0.5T : R=2T/h.
* Sweep Rate B<0.5T : decrease by a factor of 2 proportional to the field.
* Sweep Rate B<0.5T : decrease by a factor of 2 proportional to the field.





{{clear}}
{{clear}}



== Upper Magnet UMM ==
== Upper Magnet UMM ==

*I/B-Conversion: 10A=1T
*I/B-Conversion: 10A=1T



{{clear}}
{{clear}}



== Lower Magnet LMM ==
== Lower Magnet LMM ==

* I/B-Conversion: 0.8T=6.96A
* I/B-Conversion: 0.8T=6.96A



{{clear}}
{{clear}}

Revision as of 08:22, 15 December 2011

THE 9T MAGNET


... is working but this page is under construction :)

First test with 50Amps (~5.7T) worked!

In the picture a threaded rod is attracted by the magnet at 50A.

Using the Magnet  :)

  • Power Supply : OXFORD IPS 120-10 (Max.12 Amp @10Volt)
    • Software Remote-control: IPS_120-10_v03.vi
    • ! Change of Range when increasing the current at ~4.3A, when decreasing at ~3.4A! Don't get scared by the noise :)
  • Upper Magnet, Connection on window side of cryo. Power-cable are marked red (+) and blue (-).
  • Persistent current switch:
    • Upper magnet - BOX VII, X9:1/4 ,RHeater(4K)=135.
    • IHeater=55mA. The TVL-10 is used as current supply.


Magnetisation and Demagnetisation Procedure

  • After ramping the magnet the magnetized first nuclear stage must be precooled. The temperature dependence varies as T~t-1/3.
  • The first nuclear stage was demagnetized by reducing the field in steps by a factor of 2 to a final field of few mT.
  • Sweep Rate B>0.5T : R=2T/h.
  • Sweep Rate B<0.5T : decrease by a factor of 2 proportional to the field.



Upper Magnet UMM

  • I/B-Conversion: 10A=1T



Lower Magnet LMM

  • I/B-Conversion: 0.8T=6.96A


Connection

  • To Power supply: M6 (13RigschlĆ¼ssle, 6erInbus).
  • To Upper Magnet: Connection on window side.