PINBAR

From KIP Wiki
⧼kip-jumptonavigation⧽⧼kip-jumptosearch⧽

PIN BAR

  • All values are including the measuring cables!
  • Nomenclature:
    • Greek numbers = Filter Box on the top
    • Xi = Plug at the corresponding Filter Box (e.g. : VI, X3 = Plug X3 at Filter Box VI ).
    • Xi:j = Pin j at Plug Xi (e.g. : VI, X3:4 = Pin 4 at Plug X3 at Filter Box VI ).
    • R1/2 = Resistance between Pin 1 and 2 directly measured at a corresponding PIN BAR.
    • RX3:1-2 = Resistance between Pin 1 at Plug X3 of the Filter Box and Pin 2 at the PIN BAR.
    • RX3:1/2 = Resistance between Pin 1 and Pin 2 at Plug X3 of the Filter Box.


MIXING CHAMBER

There are six PIN BARS (A-F) mounted on the mixing chamber. Each consists of 16 PINS arranged in two columns. NUmber in brackets "()" corresponds to the AMP-Test-Box.


BAR A

BAR A
  • Connected with BOX IV.
  • Resistance between PIN and AMP-Test-Box R=145-148.
  • NO GROUNDS!
P CONNECTION PLG BOX COMMENT
15 nc L IV
13 nc N IV
11 nc P IV
9 nc S IV
7 nc a IV
5 nc d IV
3 nc f IV
1 nc h IV
P CONNECTION PLG BOX COMMENT
16 nc M IV
14 nc c IV
12 nc R IV
10 nc T IV
8 nc b IV
6 nc e IV
4 nc g IV
2 nc j IV

BAR B

BAR B
  • Connected with BOX III.
  • Resistance between PIN and AMP-Test-Box R=142-144.
  • NO GROUNDS!
P CONNECTION PLG BOX COMMENT
15 nc g(5) III R=150 ?
13 nc j(12) III R=800 ?
11 nc Y(6) III
9 nc Z(7) III
7 nc U(8) III
5 nc c(14) III
3 nc d(16) III
1 nc e(18) III R=170 ?
P CONNECTION PLG BOX COMMENT
16 nc f(23) III
14 nc h(11) III
12 nc X(24) III
10 nc a(25) III
8 nc T(26) III
6 nc b(13) III
4 nc V(15) III
2 nc W(17) III

BAR C

BAR C
  • Connected with BOX III.
  • Resistance between PIN and AMP-Test-Box R=141-145.
  • NO GROUNDS!
P CONNECTION PLG BOX COMMENT
15 nc K(1) III
13 nc E(30) III
11 nc N(2) III
9 T.P. 9-10 A(3) III leads through capillary
7 nc C(4) III
5 nc G(32) III
3 nc L(34) III
1 nc (36) III R=170 ?
P CONNECTION PLG BOX COMMENT
16 nc nc nc seems to be broken
14 nc F(29) III
12 nc P(20) III
10 T.P. 9-10 B(21) III capillary leads to 1K-pot
8 nc D(22) III
6 nc H(31) III
4 nc M(33) III
2 nc S(35) III

BAR D

  • supposed to be BOX III ?
BAR D
PIN CONNECTION BOX COMMENT
15 nc ?
13 nc ?
11 nc ?
9 nc ?
7 nc ?
5 nc ?
3 nc ?
1 nc ?
PIN CONNECTION BOX COMMENT
16 nc ?
14 nc ?
12 nc ?
10 nc ?
8 nc ?
6 nc ?
4 nc ?
2 nc ?


BAR E

BAR E
  • Assignment of pins at Bar E. The number of each pin corresponds with that of the AMP-BOX.
  • Resistance between Box VI and Pins: RVI-E ~ 280
  • All values are including measuring cables!
  • NO GROUNDS!
PIN CONNECTION BOX COMMENT
1 ET ST VI, X1:1
2 ET ST VI, X1:2
3 FPD VI, X2:1
4 FPD VI, X2:3
5 FPD Degauss VI, X3:1 R5/23=155.4
6 nc VI, X3:2
7 nc VI, X4:1
8 HZ(?) VI, X4:2 R8/26=1k , T.P. leads to strange heat sink at 1K
PIN CONNECTION BOX COMMENT
19 ET ST VI, X1:3
20 ET ST VI, X1:4
21 FPD VI, X2:4
22 FPD VI, X2:3
23 FPD Degauss VI, X3:4 RX3:1-4=714
24 nc VI, X3:3
25 nc VI, X4:4
26 HZ(?) VI, X4:3 RX3:1-4=1.65k


BAR F

BAR F
  • Resistance between Pin 6,7 and 8 is measured with connected Filterbox.
  • Number in brackets "()" is the corresponding AMP-TEST-BOX connection.
  • F1-F4 are connected to PIN-BAR S0 at Still-Level and via there to BOX-II.
  • NO GROUNDS!
P CONNECTION PLG BOX COMMENT
15 T.P. 15/16, ends at Still! ->n.c.! S I, X10:2 RX10:2-15=172.7
13 nc J I, X9:3 RX9:3-13=169
11 nc K I, X9:4 RX9:3-11=168.6
9 nc L,a I, X10:3 RX10:3-9=177.9
7 CT (FT) on MC, S5 at LR700 c I, X6:4, X6:3 R7/8=122
5 nc V I, X11:1 RX11:1-5=68.9
3 nc - STILL S0:4 RS0:4-3=14
1 nc b II,(10) , S0:6 RII,b-1=69
P CONNECTION PLG BOX COMMENT
16 T.P. 15/16 M,Z I, X10:4 sc with PIN 10 on backside
14 nc M,Z I, X10:4 sc with PIN 10 on backside
12 nc M,Z I, X10:4 sc with PIN 10 on backside
10 nc M,Z I, X10:4 RX10:4-10=47.7 k
8 CT (FT) on MC, S5 at LR700 A-F I, X2-X6:1-4 see here:circuit diagram
6 nc, same function as F8 A-F - R6/8=0.3, R6/7=122
4 nc - S0:1 RS0:1-4=14
2 nc Y II,(21) , S0:3 RII,Y-2=69

STILL

There are seven PIN BARS (0,A-F) mounted on the mixing chamber. Each consists of 16 PINS arranged in two columns.


BAR S:0

STILL BAR S:0
  • Connected to BOX II
  • NO GROUNDS
  • 07-11 are connected to heaters of the persistent current switch (PCS) 4P PCS on the bottom of the 4K-flange.
  • Number in brackets "()" is the corresponding AMP-TEST-BOX connection.
  • counting from top to bottom (left upper corner: 15, right bottom:2)
P CONNECTION PLG BOX COMMENT
15 MC Heater (-) F(6) II,X1:2
13 MC Heater (-) L(23) -
11 PCS 4P-2 Z(17) - RZ-11=50
9 PCS 4P-4 V(11) - RV-9=50
7 PCS com. Current X(16) - sc with 2,7,8,11, 12, 13,14,17,18,20 at AMP Test-BOX
5 Degauss cable at MC U(9) - nc at MC
3 F2 at MC Y(21) - RS0:3-F2=14
1 F4 at MC nc - no connection on backside
P CONNECTION PLG BOX COMMENT
16 MC Heater (+) E(5) II,X1:1
14 MC Heater (+) D(4) II, -
12 PCS 4P-1 B(2) II, - RB-12=50
10 PCS 4P-3 T(8) II, - RT-10=50
8 Degauss cable at MC A(1) II, - nc at MC
6 F1 at MC b(10) II, - RS0:6-F1=14
4 F3 at MC - - no connection on backside
2 nc - - no connection on backside


BAR S:A

  • BAR S:A is connected with BAR A on MC! Details will be added later.


BAR S:B

  • BAR S:B is connected with BAR B on MC! Details will be added later.


BAR S:C

  • BAR S:C is connected with BAR C on MC! Details will be added later.


BAR S:D

  • BAR S:D is connected with BAR D on MC! Details will be added later.


BAR S:E

  • BAR S:E is connected with BAR E on MC! Details will be added later.


BAR S:F

STILL BAR S:F
  • Connected to BOX V.
  • Connected with superconducting Nb-wires leading to the Cu-stage in textile tube "beta". The sc wires of tube "alpha" are directly connected with BOX V.
  • NO GROUNDS!
P CONNECTION PLG BOX COMMENT
15 nc B V R15-end=173
13 nc D V R13-end=171
11 nc F V R11-end=172
9 nc H V R9-end=140
7 nc K V R7-end=170
5 nc L V R5-end=169
3 nc P V R3-end=169
1 nc S V R1-end=168
P CONNECTION PLG BOX COMMENT
16 nc A V R16-end=172
14 nc C V R14-end=172
12 nc E V R12-end=170
10 nc G V R10-end=140
8 nc Y V R8-end=170
6 nc M V R6-end=138
4 nc N V R4-end=169
2 nc R V R2-end=169