Calibration of AuxPwr Started at 17.1.2020 at 10:30:35 Set VOL target value to 1.0 Set VOH target value to 1.2 Starting test without load Started at 17.1.2020 at 10:30:40 Firmware version is 28 Enabling monitoring Enabling all voltages, using AuxPwr's built-in enable command *** Testing voltage VZK with target value 3.0 V and tolerance +- 0.1 V ADC-value is: 3.035 V Manually measured value for VZK: 2.99 measured value for voltage VZK is inside of the tolerance range *** Testing voltage VCC with target value 9.6 V and tolerance +- 0.05 V ADC-value is: 9.74 V Manually measured value for VCC: 9.5734 measured value for voltage VCC is inside of the tolerance range *** Testing voltage DI_VBIAS with target value 1.25 V and tolerance +- 0.025 V ADC-value is: 1.266 V Manually measured value for DI_VBIAS: 1.2503 measured value for voltage DI_VBIAS is inside of the tolerance range *** Testing voltage DI_VCC33ANA with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.351 V Manually measured value for DI_VCC33ANA: 3.3023 measured value for voltage DI_VCC33ANA is inside of the tolerance range *** Testing voltage VDD12 with target value 10.65 V and tolerance +- 0.05 V ADC-value is: 10.808 V Manually measured value for VDD12: 10.6323 measured value for voltage VDD12 is inside of the tolerance range *** Testing voltage DI_VCC with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.828 V Manually measured value for DI_VCC: 1.8017 measured value for voltage DI_VCC is inside of the tolerance range *** Testing voltage VCC33INT with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.337 V Manually measured value for VCC33INT: 3.2869 measured value for voltage VCC33INT is inside of the tolerance range *** Testing voltage DI_VCCANA with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.832 V Manually measured value for DI_VCCANA: 1.8044 measured value for voltage DI_VCCANA is inside of the tolerance range *** Testing voltage VDD25 with target value 2.5 V and tolerance +- 0.05 V ADC-value is: 2.542 V Manually measured value for VDD25: 2.5014 measured value for voltage VDD25 is inside of the tolerance range *** Testing voltage VOL with target value 1.0 V and tolerance +- 0.02 V ADC-value is: 1.004 V Manually measured value for VOL: 1.0026 measured value for voltage VOL is inside of the tolerance range *** Testing voltage VDDBUS with target value 1.0 V and tolerance +- 0.02 V ADC-value is: 1.02 V Manually measured value for VDDBUS: 1.00725 measured value for voltage VDDBUS is inside of the tolerance range *** Testing voltage VOH with target value 1.2 V and tolerance +- 0.02 V ADC-value is: 1.197 V Manually measured value for VOH: 1.2007 measured value for voltage VOH is inside of the tolerance range *** Testing voltage VCC33INT_D with target value 3.9 V and tolerance +- 0.15 V ADC-value is: 4.011 V Manually measured value for VCC33INT_D: 3.9508 measured value for voltage VCC33INT_D is inside of the tolerance range *** Testing voltage VDD15TEMP with target value 15.0 V and tolerance +- 0.1 V ADC-value is: 15.224 V Manually measured value for VDD15TEMP: 14.968 measured value for voltage VDD15TEMP is inside of the tolerance range *** Testing voltage VDD5 with target value 5.5 V and tolerance +- 0.1 V ADC-value is: 5.598 V Manually measured value for VDD5: 5.5081 measured value for voltage VDD5 is inside of the tolerance range Disabling monitoring Disabling all voltages, using AuxPwr's built-in disable command