Calibration of AuxPwr Started at 28.6.2022 at 10:32:10 Responsible person for the run: LS Set VOL target value to 0.93 Set VOH target value to 1.044 Starting test without load Started at 28.6.2022 at 10:32:38 Firmware version is 33 Enabling all voltages, using AuxPwr's built-in enable command *** Testing voltage VZK with target value 3.0 V and tolerance +- 0.1 V ADC-value is: 3.029 V Manually measured value for VZK: 2.9958 measured value for voltage VZK is inside of the tolerance range *** Testing voltage VCC with target value 9.6 V and tolerance +- 0.05 V ADC-value is: 10.081 V Manually measured value for VCC: 9.8342 !!! info: measured value for voltage VCC is out of tolerance range !!! *** Testing voltage DI_VBIAS with target value 1.25 V and tolerance +- 0.025 V ADC-value is: 1.27 V Manually measured value for DI_VBIAS: 1.2577 measured value for voltage DI_VBIAS is inside of the tolerance range *** Testing voltage DI_VCC33ANA with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.339 V Manually measured value for DI_VCC33ANA: 3.3989 measured value for voltage DI_VCC33ANA is inside of the tolerance range *** Testing voltage VDD12 with target value 10.65 V and tolerance +- 0.05 V ADC-value is: 10.969 V Manually measured value for VDD12: 10.8376 !!! info: measured value for voltage VDD12 is out of tolerance range !!! *** Testing voltage DI_VCC with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.822 V Manually measured value for DI_VCC: 1.8016 measured value for voltage DI_VCC is inside of the tolerance range *** Testing voltage VCC33INT with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.34 V Manually measured value for VCC33INT: 3.3022 measured value for voltage VCC33INT is inside of the tolerance range *** Testing voltage DI_VCCANA with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.824 V Manually measured value for DI_VCCANA: 1.8023 measured value for voltage DI_VCCANA is inside of the tolerance range *** Testing voltage VDD25 with target value 2.5 V and tolerance +- 0.05 V ADC-value is: 2.537 V Manually measured value for VDD25: 2.5061 measured value for voltage VDD25 is inside of the tolerance range *** Testing voltage VOL with target value 0.93 V and tolerance +- 0.02 V ADC-value is: 0.941 V Manually measured value for VOL: 0.94795 measured value for voltage VOL is inside of the tolerance range *** Testing voltage VDDBUS with target value 1.0 V and tolerance +- 0.02 V ADC-value is: 1.011 V Manually measured value for VDDBUS: 1.01337 measured value for voltage VDDBUS is inside of the tolerance range *** Testing voltage VOH with target value 1.044 V and tolerance +- 0.02 V ADC-value is: 1.051 V Manually measured value for VOH: 1.06123 measured value for voltage VOH is inside of the tolerance range *** Testing voltage VCC33INT_D with target value 3.9 V and tolerance +- 0.15 V ADC-value is: 3.984 V Manually measured value for VCC33INT_D: 3.9415 measured value for voltage VCC33INT_D is inside of the tolerance range *** Testing voltage VDD15TEMP with target value 15.0 V and tolerance +- 0.1 V ADC-value is: 15.174 V Manually measured value for VDD15TEMP: 15.002 measured value for voltage VDD15TEMP is inside of the tolerance range *** Testing voltage VDD5 with target value 5.5 V and tolerance +- 0.1 V ADC-value is: 5.572 V Manually measured value for VDD5: 5.50073 measured value for voltage VDD5 is inside of the tolerance range Disabling monitoring Disabling all voltages, using AuxPwr's built-in disable command Starting test with load Started at 28.6.2022 at 10:33:57 connect electronic load to VDD25 and preset a load of 3.0 A calibrate lower VDD25 current value with target value 3.0 A lower ADC-value for VDD25 is 2.504 measured value for lower calib point of VDD25 is 2.998 calibrate upper current VDD25 value with target value 20.0 A now set load for VDD25 to a value of 20.0 A upper ADC-value for VDD25 is 21.612 measured value for upper calib point of VDD25 is 19.996 A Turn off VDD25 connect electronic load to VDD5 and preset a load of 0.015 A calibrate lower VDD5 current value with target value 0.015 A lower ADC-value for VDD5 is 0.018 measured value for lower calib point of VDD5 is 0.013 calibrate upper current VDD5 value with target value 0.15 A now set load for VDD5 to a value of 0.15 A upper ADC-value for VDD5 is 0.159 measured value for upper calib point of VDD5 is 0.148 A Turn off VDD5 connect electronic load to DI_VBIAS and preset a load of 0.015 A calibrate lower DI_VBIAS current value with target value 0.015 A lower ADC-value for DI_VBIAS is 0.009 measured value for lower calib point of DI_VBIAS is 0.013 calibrate upper current DI_VBIAS value with target value 0.04 A now set load for DI_VBIAS to a value of 0.04 A upper ADC-value for DI_VBIAS is 0.04 measured value for upper calib point of DI_VBIAS is 0.038 A Turn off DI_VBIAS connect electronic load to DI_VCC and preset a load of 0.015 A calibrate lower DI_VCC current value with target value 0.015 A lower ADC-value for DI_VCC is 0.0 measured value for lower calib point of DI_VCC is 0.013 calibrate upper current DI_VCC value with target value 8 A now set load for DI_VCC to a value of 8 A upper ADC-value for DI_VCC is 8.035 measured value for upper calib point of DI_VCC is 7.997 A Turn off DI_VCC connect electronic load to VOL and preset a load of 0.5 A calibrate lower VOL current value with target value 0.5 A lower ADC-value for VOL is 0.25 measured value for lower calib point of VOL is 0.497 calibrate upper current VOL value with target value 10 A now set load for VOL to a value of 10 A upper ADC-value for VOL is 9.75 measured value for upper calib point of VOL is 9.998 A Turn off VOL connect electronic load to VDDBUS and preset a load of 1 A calibrate lower VDDBUS current value with target value 1 A lower ADC-value for VDDBUS is 0.991 measured value for lower calib point of VDDBUS is 0.998 calibrate upper current VDDBUS value with target value 10 A now set load for VDDBUS to a value of 10 A upper ADC-value for VDDBUS is 9.972 measured value for upper calib point of VDDBUS is 9.998 A Turn off VDDBUS connect electronic load to DI_VCCANA and preset a load of 0.05 A calibrate lower DI_VCCANA current value with target value 0.05 A lower ADC-value for DI_VCCANA is 0.054 measured value for lower calib point of DI_VCCANA is 0.049 calibrate upper current DI_VCCANA value with target value 0.6 A now set load for DI_VCCANA to a value of 0.6 A upper ADC-value for DI_VCCANA is 0.609 measured value for upper calib point of DI_VCCANA is 0.598 A Turn off DI_VCCANA connect electronic load to VOH and preset a load of 0.5 A calibrate lower VOH current value with target value 0.5 A lower ADC-value for VOH is 0.375 measured value for lower calib point of VOH is 0.497 calibrate upper current VOH value with target value 10 A now set load for VOH to a value of 10 A upper ADC-value for VOH is 9.687 measured value for upper calib point of VOH is 9.997 A Turn off VOH connect electronic load to VDD12 and preset a load of 0.015 A calibrate lower VDD12 current value with target value 0.015 A lower ADC-value for VDD12 is 0.022 measured value for lower calib point of VDD12 is 0.013 calibrate upper current VDD12 value with target value 0.15 A now set load for VDD12 to a value of 0.15 A upper ADC-value for VDD12 is 0.153 measured value for upper calib point of VDD12 is 0.148 A Turn off VDD12 connect electronic load to DI_VCC33ANA and preset a load of 0.1 A calibrate lower DI_VCC33ANA current value with target value 0.1 A lower ADC-value for DI_VCC33ANA is 0.0 measured value for lower calib point of DI_VCC33ANA is 0.98 calibrate upper current DI_VCC33ANA value with target value 8 A now set load for DI_VCC33ANA to a value of 8 A upper ADC-value for DI_VCC33ANA is 8.085 measured value for upper calib point of DI_VCC33ANA is 7.997 A Turn off DI_VCC33ANA calibration is now finished