Calibration of AuxPwr Started at 17.1.2020 at 10:20:10 Set VOL target value to 1.0 Set VOH target value to 1.2 Starting test without load Started at 17.1.2020 at 10:20:13 Firmware version is 28 Enabling monitoring Enabling all voltages, using AuxPwr's built-in enable command *** Testing voltage VZK with target value 3.0 V and tolerance +- 0.1 V ADC-value is: 3.031 V Manually measured value for VZK: 2.9956 measured value for voltage VZK is inside of the tolerance range *** Testing voltage VCC with target value 9.6 V and tolerance +- 0.05 V ADC-value is: 9.695 V Manually measured value for VCC: 9.5775 measured value for voltage VCC is inside of the tolerance range *** Testing voltage DI_VBIAS with target value 1.25 V and tolerance +- 0.025 V ADC-value is: 1.263 V Manually measured value for DI_VBIAS: 1.2499 measured value for voltage DI_VBIAS is inside of the tolerance range *** Testing voltage DI_VCC33ANA with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.337 V Manually measured value for DI_VCC33ANA: 3.2991 measured value for voltage DI_VCC33ANA is inside of the tolerance range *** Testing voltage VDD12 with target value 10.65 V and tolerance +- 0.05 V ADC-value is: 10.969 V Manually measured value for VDD12: 10.8335 measured value for voltage VDD12 is out of tolerance range *** Testing voltage DI_VCC with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.824 V Manually measured value for DI_VCC: 1.8019 measured value for voltage DI_VCC is inside of the tolerance range *** Testing voltage VCC33INT with target value 3.3 V and tolerance +- 0.1 V ADC-value is: 3.34 V Manually measured value for VCC33INT: 3.3018 measured value for voltage VCC33INT is inside of the tolerance range *** Testing voltage DI_VCCANA with target value 1.8 V and tolerance +- 0.03 V ADC-value is: 1.824 V Manually measured value for DI_VCCANA: 1.8019 measured value for voltage DI_VCCANA is inside of the tolerance range *** Testing voltage VDD25 with target value 2.5 V and tolerance +- 0.05 V ADC-value is: 2.537 V Manually measured value for VDD25: 2.5047 measured value for voltage VDD25 is inside of the tolerance range *** Testing voltage VOL with target value 1.0 V and tolerance +- 0.02 V ADC-value is: 1.019 V Manually measured value for VOL: 1.00323 measured value for voltage VOL is inside of the tolerance range *** Testing voltage VDDBUS with target value 1.0 V and tolerance +- 0.02 V ADC-value is: 1.011 V Manually measured value for VDDBUS: 1.00168 measured value for voltage VDDBUS is inside of the tolerance range *** Testing voltage VOH with target value 1.2 V and tolerance +- 0.02 V ADC-value is: 1.47 V Manually measured value for VOH: 1.2052 measured value for voltage VOH is inside of the tolerance range *** Testing voltage VCC33INT_D with target value 3.9 V and tolerance +- 0.15 V ADC-value is: 3.986 V Manually measured value for VCC33INT_D: 3.9418 measured value for voltage VCC33INT_D is inside of the tolerance range *** Testing voltage VDD15TEMP with target value 15.0 V and tolerance +- 0.1 V ADC-value is: 15.195 V Manually measured value for VDD15TEMP: 15.002 measured value for voltage VDD15TEMP is inside of the tolerance range *** Testing voltage VDD5 with target value 5.5 V and tolerance +- 0.1 V ADC-value is: 5.568 V Manually measured value for VDD5: 5.5001 measured value for voltage VDD5 is inside of the tolerance range Disabling monitoring Disabling all voltages, using AuxPwr's built-in disable command